- Home
- Fetal Monitor
- 3-Parameter Fetal Monitor for Single and Twin Pregnancies
Size
295*240*73 (mm) (11.6″ *9.4″ *2.9″)
Weight
1.75kg/3.9lb
Display
800×480, 7″ color touchscreen
Power Supply
AC 100-240V, 50/60Hz
Power Consumption
< 70 VA
Battery
14.8V lithium battery, 2200mAh
Device Storage
3000 hours
Charging Time
3 hours
Battery Working Time
5 hours
Language
Support 9 language displays: English, Chinese, Turkish, Portuguese, Russian, French, Spanish, Polish, Ukrainian
Admin Password
26833601
Transducer
9-crystal wide beam pulsed doppler
Strength
<5mW/cm2
Working Frequency
1.0MHz
Signal Processing
special digital signal
Measurement Range
50~210 bpm / 30-240bpm
Alarm Range
High limit: 160,170,180,190 bpm
Lower limit: 90,100,110,120 bpm
Maximum audio output: 2 W
Working Environment
Temperature: +5℃ ~ +40℃; Humidity: < 30%~75%
Atmospheric pressure: 86kPa ~ 106kPa
Transport and Storage Temperature
Temperature: -10℃ ~ +55℃; Humidity: < 95%
Atmospheric pressure: 86kPa ~ 106kPa
Probe Acoustic Output
In accordance with the provisions of 1992 IEC1157, negative peak sound pressure shouldn’t exceed 1 MPa, and the beam intensity shouldn’t exceed 20mW/cm2. Spatial peak instantaneous average intensity density shouldn’t exceed 100mW/cm2. The sound intensity of this model does not exceed 5mW/cm2. Sound output meets the conditions exempted from publication.
Alarm Indicator
Flashing red: High level alarm, flash frequency: 2Hz
Flashing Yellow
Intermediate level alarm, flash frequency: 0.5Hz
Constant Yellow
Low level alarm: Keep on flashing