- Home
- Patient Monitor
- 12-Inch Patient Monitor
WEIGHT
6.6lb (Rechargeable lithium battery included)
SIZE
12.6″×6.3″×10.6″
DISPLAY
12.1″ color TFT-LCD
POWER
Power supply: 100-240VAC, 50/60Hz
Battery capacity: 5000mAh (3 hours)
WARRANTY
1 year (Service Life: 5 years)
CONDITION
Brand new
FDA APPROVED
Yes
CE CERTIFICATED
Yes
DATA STORAGE
Trend data: 480hrs
Arrhythmia alarm review: 60 8-sec events
NIBP review: 4800 groups
NETWORK CONNECTION
Wired LAN, RJ45 port
DRUG DOSE CALCULATION
15 drug types
WORKING MODE
Monitoring mode; Demo mode
SUITABLE FOR
Adults, Pediatrics, Neonates
Lead mode
3-lead or 5-lead
Lead selection
I, II, III, aVR, aVL, aVF, V
Waveform
5-lead, 2-channel; 3-lead, 1-channel
Gain
2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV
Weight
250g
Scan speed
12.5mm/s, 25 mm/s, 50 mm/s
HR:
Measurement & alarm range:
15~350bpm
Accuracy
±1% or ±1bpm, whichever is greater
Alarm accuracy:
± 2 bpm
ST-segment monitoring:
Measurement & alarm range:
-2.0mV~+2.0mV
Accuracy:
-0.8mv~+0.8mv ±0.04mv or ±10%, whichever is greater
Arrhythmia analysis:
ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Defib-proof & ESU-proof:
Yes (with our special ECG cable)
Pacemaker detection:
Yes
Method:
Oscillometry
Mode:
Manual/AUTO/Continuous
Measurement interval in AUTO mode:
1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes
Measurement period in Continuous mode:
5 minutes
Measurement and alarm range:
10 ~ 270 mmHg
Resolution:
1 mmHg
Cuff size:
25-35cm
Cuff pressure accuracy:
±3 mmHg
Measurement accuracy:
Maximal mean deviation:
±5 mmHg
Maximal standard deviation:
8mmHg
Over-pressure protection:
Adult mode:
297±3 mmHg
Pediatric mode:
240±3 mmHg
Neonatal mode:
147±3 mmHg
Method:
R-F(RA-LL) Impedance
Measurement and alarm range:
0~150rpm
Resolution:
1 rpm
Measurement accuracy:
±2 rpm
Apnea alarm:
10~40s
Scan speed:
6.25 mm/s, 12.5 mm/s, 25 mm/s
Measurement and alarm range:
0 ~ 100%
Resolution:
1%
Measurement accuracy:
70%~100%: ±2%; 0%~69%: unspecified
Channel:
dual-channel
Measurement and alarm range:
0 ~ 122°F
Resolution:
0.1 °F
Accuracy:
±0.1 °F
Measurement and alarm range:
0 ~ 100%
Resolution:
1%
Measurement accuracy:
70%~100%: ±2%; 0%~69%: unspecified